1D non-refocused 1H-coupled 15N spectrum
(using INEPT)

PRELIMINARY SET-UP PARAMETER SET
a conventional 1H-decoupled 15N spectrum with rpar N15INEPT all . Change the pulse program to ineptnd
MODIFY SPECIFIC PARAMETERS

By default, the following parameters are set to: ns=32, ds=2, d1=5, sw=519 and td=32k. Change them if required. The interpulse delay d4 is optimized to 1/(4*1JNH). In the case of formamide this value is set to 2.8 ms.

ACQUISITION
PROCESSING
PLOT
OBSERVATIONS

(au_zg)) (proc_1d).

OTHER RELATED VERSIONS
This same experiment can be recorded without refocusing and without decoupling. Only the pulse program might be changed: 1H-decoupled using INEPT (ineptrd)
Refocused 1H-coupled using INEPT (ineptpnd)