SPECTRA
 

Sample: Urea/Methanol
Solvent: DMSO
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: High-power 15N decouple pulse calibration
Pulse program: decp90f3 ( Pulse Diagram)
Experiment Time: 5 minute
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=16k
o1p=4
o2p=76
sw=8
ns=1
ds=0
d1=10
d2=5.649m
ef using lb=0.3
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. pl2 is set to -3dB and p3 is varied from 12 to 26 microseconds using a 2 microsecond step. When using a small p3 value (for instance, 12u), we observe antiphase signals at 5.4 ppm. P3 is increased until these signals become zero (near 18u) and this is the 90 degree 15N pulse from the decoupler at high-power level.
A similar procedure is followed to measure the
Low-power 15N pulse calibration from decoupler (for GARP).