SPECTRA
 

Sample: Urea/Methanol
Solvent: DMSO
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: Low-power 15N decouple pulse calibration
Pulse program: decp90f3 ( Pulse Diagram)
Experiment Time: 5 minute
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=16k
o1p=4
o2p=76
sw=8
ns=1
ds=0
d1=10
d2=5.649m
ef using lb=0.3
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. p3 is set to 120u and pl2 is varied from 11dB to 18dB using a 1dB step. When using a small pl2 value (for instance, 11dB), we observe antiphase signals at 5.4 ppm. Pl2 is increased until these signals become zero (near 15dB) and this is the 90 degree 15N pulse from the decoupler at low-power level to be used for GARP decoupling.
A similar procedure is followed to measure the
High-power 15N pulse calibration from decoupler.
RELATED ITEMS
See list of spectra of Urea/Methanol sample
OBSERVATIONS
 
PARAMETER SET FILE