SPECTRA
 

Sample: Urea/Methanol
Solvent: DMSO
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: High-power 13C decouple pulse calibration
Pulse program: decp90 ( Pulse Diagram)
Experiment Time: 5 minute
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=16k
o1p=4
o2p=49.5
sw=8
ns=1
ds=0
d1=10
d2=3.59m
ef using lb=0.3
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. pl2 is set to 0dB and p3 is varied from 8 to 17 microseconds using a 1 microsecond step. When using a small p3 value (for instance, 8u), we observe antiphase signals at 3.2 ppm. P3 is increased until the signals become zero (near 12u) and this is the 90 degree 13C pulse from the decoupler at high-power level.
A similar procedure is followed to measure the
Low-power 13C pulse calibration from decoupler (for GARP).
RELATED ITEMS
See list of spectra of Urea/Methanol sample
OBSERVATIONS
 
PARAMETER SET FILE