Sample: Urea/Methanol Solvent: DMSO Spectrometer: AVANCE 500 Probehead: Inverse Broadband with z-Gradients Experiment: High-power 13C decouple pulse calibration Pulse program: decp90 ( Pulse Diagram) Experiment Time: 5 minute Details: AVANCE Tutorial |
ACQUISITION | PROCESSING |
td=16k
o1p=4 o2p=49.5 sw=8 ns=1 ds=0 d1=10 d2=3.59m |
ef using lb=0.3
si=16k |
This pulse calibration is performed using the paropt AU program. pl2 is set to 0dB and p3 is varied from 8 to 17 microseconds using a 1 microsecond step. When using a small p3 value (for instance, 8u), we observe antiphase signals at 3.2 ppm. P3 is increased until the signals become zero (near 12u) and this is the 90 degree 13C pulse from the decoupler at high-power level.RELATED ITEMSA similar procedure is followed to measure the Low-power 13C pulse calibration from decoupler (for GARP).
See list of spectra of Urea/Methanol sampleOBSERVATIONS