SPECTRA
 

Sample: Urea/Methanol
Solvent: DMSO
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: Low-power 13C decouple pulse calibration
Pulse program: decp90 ( Pulse Diagram)
Experiment Time: 5 minutes
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=16k
o1p=4
o2p=49.5
sw=8
ns=1
ds=0
d1=10
d2=3.59m
ef using lb=0.3
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. p3 is set to 80u and pl2 is varied from 13 to 20 dB using a 1 dB step. When using a small pl2 value (for instance, 13u), we observe antiphase signals at 3.2 ppm. pl2 is increased until the signals become zero (near 16dB) and this is the 90 degree 13C pulse from the decoupler at low-power level. These values are used for 13C GARP decoupling in all inverse applications.
A similar procedure is followed to measure the
High-power 13C pulse calibration from decoupler.
RELATED ITEMS
See list of spectra of Urea/Methanol sample
OBSERVATIONS
 
PARAMETER SET FILE