Sample: 1,2-dibromopropionic acid
Solvent: CDCl3
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: Selective 1H observe pulse calibration
Pulse program: selzg ( Pulse Diagram)
Experiment Time: 4 minute
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=32k
o1p=0
sw=10
ns=1
ds=0
d1=10
spnam1=gauss1.1000
ef using lb=0.3
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. o1 is set on resonance of the selected resonance, p11 is set to 80m and sp1 is varied from 75 to 61dB using a -1dB step. The maximum signal stands for the selective 90 degree pulse (in our case, around 70-72dB) that is confirmed by the first null point (selective 180 degree pulse) when the pulse power level is decreased 6dB (64dB). Thus, as an example, we can use the following p11/sp1 pair values for a 1% truncated gaussian shaped pulse as a function of the required selectivity:
  • For a selective 90 pulse:
    • 80ms a 70 dB
    • 40ms a 64 dB
    • 20ms a 58 dB
  • For a selective 180 pulse:
    • 80ms a 64 dB
    • 40ms a 58 dB
    • 20ms a 52 dB
  • For a selective 270 pulse:
    • 90ms a 61 dB
    • 60ms a 58 dB
    • 30ms a 52 dB
  • RELATED ITEMS
    See list of spectra of 1,2-dibromopropionic acid
    OBSERVATIONS
     
    PARAMETER SET FILE
    Start from the standard SELZG1H parameter set