|
PARAMETER SETRecord a conventional 1H spectrum. Note the o1 of the selected resonance.
a selective excitation with rpar SELZG1H all (the pulse program selzgMODIFY SPECIFIC PARAMETERS
The most important parameters to modify are the shape, the duration and the corresponding power level of the selective pulse (p11). The shape must be created from the xShape tool (for instance, spnam1=gauss1024) and the power level (sp1) and the selective pulse duration (p11) must be calibrated as described in selective pulse calibration using selzg.ACQUISITIONo1 is set on-resonance of a well-isolated selected resonance.
(minimum, ns=4 and ds=2).PROCESSING
PLOT
OBSERVATIONS
We can use other selective pulse lengths and power levels as a function of the required selectivity. Each 6 dB, the pulse duration is doubled and each 20 dB, the pulse duration is multiplied by 10. On the other hand, the duration of the corresponding selective 180º pulse is the double of the calculated selective 90º pulse at the same power level. The following table can be used as a reference:OTHER RELATED VERSIONS
10 ms 20 ms 40 ms 80 ms 100 ms selective 90º 61 67 73 79 81 selective 180º 55 61 67 73 75 selective 270º 52 58 64 70 72 Once the calibration has been succesfully carried out, the parameter set can be stored with the wpar command in order to be used in the future (for instance, wpar selzg all).
When applying selective 90 or 270 degree pulses in other sequences, there is need for phase difference calculation between power levels sp1 and pl1. This phase correction must be applied on the selective pulses.