PARAMETER SET
a 2D 1H-1H TOCSY spectrum (using MLEV-17) with (the pulse program mlevphMODIFY SPECIFIC PARAMETERS
It is generally recommended that TOCSY experiments be run without sample spinning.ACQUISITION
The power level (pl10) of this mixing time will be set according to the Tutorial: 90º 1H observe pulse calibration.
Finally, set the number of scans as a function of the sample concentration (by default, ns 8 and ds 4).
PROCESSING
PLOTPhase correct the spectrum.
OBSERVATIONS
OTHER RELATED VERSIONS(au_getlcosy)) (proc_2dhom).
Alternatively, phase-sensitive 2D TOCSY experiments using a z-filtered DIPSI-2 mixing process is also available.
List of Related Experiments:
Phase-cycle TOCSY experiments
Gradient-enhanced TOCSY experiments
Gradient-enhanced TOCSY experiments for H2O