PARAMETER SETRecord a conventional 1H spectrum. Note the o1 of the selected resonance. Perform the selective excitation on this selected proton resonance using a selective 270 degree pulse.
a Selective 1D NOESY spectrum with rpar SELNO1H all (the pulse program selnoMODIFY SPECIFIC PARAMETERS
ACQUISITIONIf using a selective 270 degree pulse (p11), d14 is fixed to a minimum value of 20u. When using a selective 90 degree pulse (p11), this delay is optimized to (p11)/2 + d14 ~ 1/JHH
The duration of the NOESY mixing time which is defined by a vd-list (l4 is the number of delays in this list). On the other hand, a fixed mixing time value can be specified.
A minimum eigth scan cycle is needed (ns=8 and ds=4).
PROCESSING
PLOT
OBSERVATIONS OTHER RELATED VERSIONS
However, improved results can be obtained using the gradient-based version (see selective ge-1D NOESY experiment).