Sample: CHCl3 Solvent: Acetone-d6 Spectrometer: AVANCE 500 Probehead: Inverse Broadband with z-Gradients Experiment: 1H lineshape and hump test Pulse program: zg30 ( Pulse Diagram) Experiment Time: 2 minutes Details: AVANCE Tutorial |
ACQUISITION | PROCESSING |
td=32k
o1p=7.7 swh=1000 ns=8 ds=0 d1=2 |
ft using si=16k |
Spinning sidebands are specified with 8 transients and must be below 1% for all standard 5mm and 8mm probes.RELATED ITEMSThe lineshape test is also performed without sample rotation on all probes with watersuppression capability.
![]()
See list of spectra of CDCl3OBSERVATIONS
PARAMETER SET FILE
Start from the standard PROHUMP parameter set
Written by Teodor Parella Copyright © 1998-2008 BRUKER Biospin. All rights reserved. |