SPECTRA
 

Sample: CHCl3
Solvent: Acetone-d6
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: 1H lineshape and hump test
Pulse program: zg30 ( Pulse Diagram)
Experiment Time: 2 minutes
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=32k
o1p=7.7
swh=1000
ns=8
ds=0
d1=2
ft using si=16k
COMMENTS
Spinning sidebands are specified with 8 transients and must be below 1% for all standard 5mm and 8mm probes.

The lineshape test is also performed without sample rotation on all probes with watersuppression capability.

RELATED ITEMS
See list of spectra of CDCl3
OBSERVATIONS
 
PARAMETER SET FILE
Start from the standard PROHUMP parameter set