SPECTRA
 

Sample: ASTM
Solvent: C6D6
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: Observe 1H pulse calibration
Pulse program: zg ( Pulse Diagram)
Experiment Time: 10 minutes
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=32k
o1p=67
swh=3000
ns=1
ds=0
d1=10
ef using lb=3.5
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. pl1 is set to 0dB and p1 is varied from 1 to 16 microseconds using a 1 microsecond step. The first maximum signal stands for the 90 degree pulse that is confirmed by the first null point when the pulse length is doubled. Also note that wrong calibration is sometimes performed when using proton decoupling.
RELATED ITEMS
See list of spectra of ASTM
OBSERVATIONS
 
PARAMETER SET FILE