Sample: ASTM Solvent: C6D6 Spectrometer: AVANCE 500 Probehead: Inverse Broadband with z-Gradients Experiment: Observe 1H pulse calibration Pulse program: zg ( Pulse Diagram) Experiment Time: 10 minutes Details: AVANCE Tutorial |
ACQUISITION | PROCESSING |
td=32k
o1p=67 swh=3000 ns=1 ds=0 d1=10 |
ef using lb=3.5
si=16k |
This pulse calibration is performed using the paropt AU program. pl1 is set to 0dB and p1 is varied from 1 to 16 microseconds using a 1 microsecond step. The first maximum signal stands for the 90 degree pulse that is confirmed by the first null point when the pulse length is doubled. Also note that wrong calibration is sometimes performed when using proton decoupling.RELATED ITEMS
See list of spectra of ASTMOBSERVATIONS
PARAMETER SET FILE