Magnitude-mode ge-2D 1H-13C HMBC

PRELIMINARY SET-UP
  • PARAMETER SET
    a phase-sensitive ge-2D 1H-13C HMBC spectrum with (the pulse program hmbcgpndqf
    MODIFY SPECIFIC PARAMETERS
    HMBC

    Standard gradient parameters:

  • 50%
  • 30%
  • 40.1%
  • Set the number of scans as a function of the sample concentration (the minimum required is ns 2 and ds 16).

    By default, the evolution period is set to d6=65 ms.

  • ACQUISITION
    PROCESSING

    PLOT
    OBSERVATIONS

    (au_getlinv)) (proc_2dpl).

    OTHER RELATED VERSIONS
    The HMBC experiment can be also recorded using a low-pass J filter to minimize undesired direct responses.

    List of Related Experiments:

    Phase-cycle HMBC experiments
    Gradient-enhanced HMBC experiments