PARAMETER SETRecord a conventional 1H spectrum. Note the o1 of the selected resonance. Perform the selective excitation on this selected proton resonance.
a selective 1H COSY spectrum with rpar SELCO1H all (the pulse program selcoMODIFY SPECIFIC PARAMETERS
ACQUISITIONThe delay for homonuclear coupling evolution is optimized to (p11)/2 + d14 ~ 1/(2*JHH) where p11 is the length of the selective 90 pulse and d14 must be fixed.
A minimum eigth-scan cycle is needed (ns=8 and ds=4).
PROCESSING
Transform with and corrects the phase (peaks must present an anti-phase pattern).PLOT
OBSERVATIONS
OTHER RELATED VERSIONS
A similar experiment is available to record selective 1D X-X COSY experiments using 1H broadband decoupling through all pulse sequence (see selcopg)