Go to eNMR
1D INADEQUATE

PRELIMINARY SET-UP PARAMETER SET
From the reference 13C spectrum create a new dataset with and change the pulse program to pulprog inad1d. All other parameters as in the 13C spectrum. 
MODIFY SPECIFIC PARAMETERS
It is very important to use a long recycle delay, usually d1=5*T1(13C) (for protonated carbons, start with d1=3 s).

The interpulse delay (d4) is optimized to 1/[4*JCC] (d4=7.6 msec, calculated from 1JCC=33 Hz).

Set ns as a function of sample concentration (minimum phase cycle with ns=16 and ds=16). Due to its low sensitivity, set ns to a large value (ns=1024)

ACQUISITION
PROCESSING
Adjust the antiphase pattern of the 13C resonances.
PLOT
OBSERVATIONS
OTHER RELATED VERSIONS
Several versions are also available:
1D INADEQUATE experiment using composite pulses (pulprog inadcp1d).
1D refocused INADEQUATE experiment (pulprog inadrd1d).