13C lineshape and hump test

PRELIMINARY SET-UP
Standard line shape test sample: 60% benzene/40% p-dioxane.
  • Record a conventional 1H spectrum.
  • Record a conventional 13C spectrum.
  • PARAMETER SET
    The calibration of the 13C line shape is performed using the standard C13HUMP parameter set (the pulse program is set to pulprog zgcw30). The width of the benzene resonance line is measured at the height of 0.55% and 0.11%. 

    MODIFY SPECIFIC PARAMETERS
    The test is usually performed using a sample spinning rate of 20 Hz.

    o1 and o2 are set on resonance of the 13C and 1H benzene resonances, respectively.

    Set p1 to the 90º 13C observe pulse and pl14 to the low power level for 1H decoupler pulse decoupler attenuation for continuous wave decoupling.

    ACQUISITION
    PROCESSING
    . Reference the benzene line to 0.0 Hz.
    PLOT
    Plot from 7 to -8 Hz with a CX=15 cm (1.0 Hz/cm) and a CY=1000 cm. Determine the lineshape at 5.5 cm and 1.1 cm height. The plot window must also show the spinning sidebands.
    OBSERVATIONS