15N sensitivity test

PRELIMINARY SET-UP
Standard 15N sensitivity test sample: 90% formamide in DMSO-d6.
  • Record a 1H-decoupled 15N spectrum.
  • PARAMETER SET
    The calibration of the 15N sensitivity test is performed using the standard 1H-decoupled 15N parameter set (the pulse program is set to pulprog zgig). This test is performed using the inverse gated decoupling scheme in order to avoid the build up of NOE. 

    MODIFY SPECIFIC PARAMETERS
    It is very important to use a long recycle delay, usually d1=5*T1(15N) (for instance d1=500 s). Use a single scan (ns=1).

    Set p1 to the 90º 15N observe pulse at high power level pl1. This pulse is calibrated reaching the maximum signal for 90º and null point in 180º.

    ACQUISITION
    PROCESSING
    . Reference the line to 0 ppm and plot the full spectrum.
    PLOT
     Plot the noise with a minimum amplitude of 1.5 cm over a minimum of 100 Hz.
    OBSERVATIONS