SPECTRA
 

Sample: Triphenylphosphate (TPP)
Solvent: Acetone-d6
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: 31P observe pulse calibration
Pulse program: zg ( Pulse Diagram)
Experiment Time: 4 minute
Details: AVANCE Tutorial
Acquisition: Test Wizard
PARAMETERS
ACQUISITION PROCESSING
td=32k
o1p=-16
sw=50
ns=1
ds=0
d1=10
ef using lb=5
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. pl1 is set to 0dB and p1 is varied from 8 to 26 microseconds using a 2 microsecond step. The first maximum signal stands for the 90 degree pulse (in our case, about 10-11u) that is confirmed by the first null point when the pulse length is doubled (22u).