SPECTRA
 

Sample: EthylBenzene
Solvent: CDCl3
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: 1H observe pulse calibration
Pulse program: zg ( Pulse Diagram)
Experiment Time: 4 minute
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=32k
o1p=4
sw=10
ns=1
ds=0
d1=10
ef using lb=0.3
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. pl1 is set to 0dB and p1 is varied from 5 to 24 microseconds using a 1 microsecond step. The first maximum signal stands for the 90 degree pulse (in our case, 10u) that is confirmed by the first null point when the pulse length is doubled (20u).
RELATED ITEMS
See list of spectra of EthylBenzene
OBSERVATIONS
 
PARAMETER SET FILE
Start from the standard PROTON parameter set