SPECTRA
 

Sample: D2O
Solvent: CDCl3
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: Gradient Recovery Test
Pulse program: preempgp2 ( Pulse Diagram)
Experiment Time: 3 minutes
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=16k
o1p=4.3
sw=10
p16=1.5m
gpz1=50%
gpnam1=square.100
ns=1
ds=0
d1=5
ef using lb=0.3
si=16k
COMMENTS
This calibration is performed using the AU paropt program. p16 is set to 1.5m and d16 is varied from 10 to 300 microseconds using a 10 microsecond step. Complete signal recovery is achieved using a delay of 20 usec.
RELATED ITEMS
See list of spectra of D2O
OBSERVATIONS
 
PARAMETER SET FILE