Sample: D2O Solvent: CDCl3 Spectrometer: AVANCE 500 Probehead: Inverse Broadband with z-Gradients Experiment: Gradient Recovery Test Pulse program: preempgp2 ( Pulse Diagram) Experiment Time: 3 minutes Details: AVANCE Tutorial |
ACQUISITION | PROCESSING |
td=16k
o1p=4.3 sw=10 p16=1.5m gpz1=50% gpnam1=square.100 ns=1 ds=0 d1=5 |
ef using lb=0.3
si=16k |
This calibration is performed using the AU paropt program. p16 is set to 1.5m and d16 is varied from 10 to 300 microseconds using a 10 microsecond step. Complete signal recovery is achieved using a delay of 20 usec.RELATED ITEMS