SPECTRA
 

Sample: ASTM
Solvent: C6D6
Spectrometer: AVANCE 500
Probehead: Inverse Broadband with z-Gradients

Experiment: Low-power Decoupler 1H pulse calibration
Pulse program: decp90 ( Pulse Diagram)
Experiment Time: 10 minutes
Details: AVANCE Tutorial
PARAMETERS
ACQUISITION PROCESSING
td=32k
o1p=67
o2p=3.7
swh=3000
ns=1
ds=0
d1=30
ef using lb=3.5
si=16k
COMMENTS
This pulse calibration is performed using the paropt AU program. p3 is set to 80 microseconds and pl2 is varied from 15 to 25 dB using a 1 dB step. With pl2=15dB, we observe an antiphase triplet, with the two outer lines in phase and the center line in opposite phase. pl12 is increased until this triplet becomes nul (in our case , 20dB). An analog procedure is followed for the High-power 1H Decoupler Pulse Calibration (for CPD)