Sample: ASTM Solvent: C6D6 Spectrometer: AVANCE 500 Probehead: Inverse Broadband with z-Gradients Experiment: Low-power Decoupler 1H pulse calibration Pulse program: decp90 ( Pulse Diagram) Experiment Time: 10 minutes Details: AVANCE Tutorial |
ACQUISITION | PROCESSING |
td=32k
o1p=67 o2p=3.7 swh=3000 ns=1 ds=0 d1=30 |
ef using lb=3.5
si=16k |
This pulse calibration is performed using the paropt AU program. p3 is set to 80 microseconds and pl2 is varied from 15 to 25 dB using a 1 dB step. With pl2=15dB, we observe an antiphase triplet, with the two outer lines in phase and the center line in opposite phase. pl12 is increased until this triplet becomes nul (in our case , 20dB). An analog procedure is followed for the High-power 1H Decoupler Pulse Calibration (for CPD)RELATED ITEMS