1D water presaturation

PRELIMINARY SET-UP PARAMETER SET
From this reference 1H spectrum create a new dataset with edc and change the pulse program (pulprog zgpr). All other parameters as in the 1H spectrum. 

Otherwhise, a standard parameter set (ZGPR) is also available. a conventional 1H presaturated spectrum with rpar ZGPR all (the pulse program zgpr

MODIFY SPECIFIC PARAMETERS
Presaturation experiments are recorded without sample spinning.
Set o1 on-resonance to the solvent resonance to be presaturated (o1 may be finely adjusted in gs mode).

The power level and the duration of the presaturation are defined by  pl19 (start with 55 dB) and d1 (2 s), respectively.

ACQUISITION
(minimum number of scans ns=8).
PROCESSING
PLOT
OBSERVATIONS

(au_watersc)) (proc_1d).

OTHER RELATED VERSIONS
 Solvent suppression can be achieved by other methods.