From this reference 1H spectrum create a new dataset with edc and change the pulse program (pulprog zgpr). All other parameters as in the 1H spectrum.MODIFY SPECIFIC PARAMETERSOtherwhise, a standard parameter set (ZGPR) is also available. a conventional 1H presaturated spectrum with rpar ZGPR all (the pulse program zgpr
Presaturation experiments are recorded without sample spinning.
Set o1 on-resonance to the solvent resonance to be presaturated (o1 may be finely adjusted in gs mode).ACQUISITIONThe power level and the duration of the presaturation are defined by pl19 (start with 55 dB) and d1 (2 s), respectively.
(minimum number of scans ns=8).PROCESSING PLOT OBSERVATIONS
OTHER RELATED VERSIONS(au_watersc)) (proc_1d).
Solvent suppression can be achieved by other methods.