ge-1D 1H-13C HMQC spectrum with refocusing but not decoupling

PRELIMINARY SET-UP PARAMETER SET
a conventional 1D 1H-13C HMQC spectra with rpar HMQC1D all and change the pulse program to hmqcgpnd1d that
MODIFY SPECIFIC PARAMETERS

d2 is the evolution delay d2 is optimized to 1/(2*1JCH) and, by default, set to 2.34 ms (for CHCl3).

Finally, gradient parameters must be defined. Standard values can be:

gpnam1=gpnam2=gpnam3=SINE.100
gpz1=50
gpz2=30
gpz3=40.1
p16=1ms
d16=100u

On the other hand, the following standard acquisition parameters can be modified as usual: d1, ns, ds, sw and td.

ACQUISITION
PROCESSING
PLOT
OBSERVATIONS

(au_zg)) (proc_1d).