Selective 90º 13C Observe Pulse Calibration

The selective 90º 13C observe pulse can be calibrated from the high-power 90º 13C observe pulse calibration parameter set-up. Proceed as follows:
  • Change the pulse program to selzg
  • Define the desired pulse shape in spnam1
  • Set o1 on the selected resonance
  • Define the pulse duration p11
  • Reach the maximum signal intensity by varying the sp1 power level