The selective 90º 13C observe pulse can be calibrated from the high-power 90º 13C observe pulse calibration parameter set-up. Proceed as follows: Change the pulse program to selzg Define the desired pulse shape in spnam1 Set o1 on the selected resonance Define the pulse duration p11 Reach the maximum signal intensity by varying the sp1 power level