Shaped 90º 13C Decoupler Pulse Calibration

The 90º 13C decoupler pulse using a shaped pulse can be calibrated from the high-power 90º 13C decoupler pulse calibration parameter set-up. Proceed as follows:
  • Change the pulse program to decp90sp
  • Define the desired pulse shape in spnam2
  • Set o2 on the selected resonance
  • Define the pulse duration p13
  • Reach the nulling signal intensity by varying the sp2 power level