The 90º 13C decoupler pulse using a shaped pulse can be calibrated from the high-power 90º 13C decoupler pulse calibration parameter set-up. Proceed as follows: Change the pulse program to decp90sp Define the desired pulse shape in spnam2 Set o2 on the selected resonance Define the pulse duration p13 Reach the nulling signal intensity by varying the sp2 power level