a gradient-based sensitivity-enhanced 3D HBHA(CO)NH experiment with (the pulse program hbhaconhgp3dMODIFY SPECIFIC PARAMETERS
Channel f1 f2 f3 Nucleus 1H 13C 15N Offset o1p o2p o3p Value 4.75 ppm 37.36 ppm 115.5 ppm
The following hard and shaped pulses applied on the three channels must be previously calibrated:ACQUISITION PROCESSING
- Hard 90º and 180º 1H pulses via the observe f1 channel (p1 and p2 at pl1).
- 90º degree 1H pulse for broadband DIPSI-2 decoupling via the observe f1 channel (pcpd1 at pl19 using cpdprg1=dipsi2).
- Selective 90 13C pulses applied on the CA/CB or on the CO resonances via the decoupler f2 channel (p13 (about 409u) at sp2, using spnam2=G4.256, spoff2=0).This same pulse is also applied to perform time reversed selective 90 pulses (p13 (about 409u) at sp8, using spnam8=G4tr.256, spoff8=0).
- Selective 180 13C pulses applied on the CA/CB or on the CO resonances via the decoupler f2 channel (p14 (about 256u) at sp3, using spnam3=G3.256, spoff3=0).
- Selective 180 13C pulses applied off-resonance on the CO resonances via the decoupler f2 channel (p14 (about 256u) at sp5, using spnam5=G3.256, spoff5=+130 ppm).
- Selective 180 13C pulses applied off-resonance on the CA resonances via the decoupler f2 channel (p14 (about 256u) at sp7, using spnam7=G3.256, spoff7=-130 ppm).
- Selective off-resonance CA decoupling via the decoupler f2 channel (pcpd2 at pl12=sp15 using cpdprg2=mlevsp180, spnam15=G3.256, and spoff15=-120ppm)
- Hard 90º and 180º 15N pulses via the decoupler f3 channel (p21 and p22 at pl3).
- 90º 15N pulse for GARP decoupling during acquisition via the decoupler f3 channel (pcpd3 at pl16 using cpdprg3=garp).
Otherwhise, all required acquisition parameters can be displayed with the ased command. By default:
Dimension F3 F2 F1 Nucleus 1H 15N 1H TD 2048 40 128 Spectral width NH (14.08 ppm) N (32 ppm) 1HA+1HB (14.08 ppm) Interpulse delays:
d4 is optimized to 1/(4*JC-H) (1.8 ms)
d21 and d27 are optimized to 1/(4*JN-CO) (12.4 ms)
d22 and d23 are optimized to 1/(4*JC-C) (3.6 ms)
d26 is optimized to 1/(4*JN-H) (2.3 ms)All other delays are automatically calculated from these values. Gradient parameters are already defined and the user must be set the number of scans and dummy scans (by default, ns=4 and ds=16) as a function of sample concentration. The standard gradient parameters are:
Gradient Number Strength Duration Shape 1 gpz1=3 p16=500u gpnam1=sine.100 2 gpz2=2 p19=1m p30=1.5m gpnam2=sine.100 3 gpz3=30 p19=1m gpnam3=sine.100 4 gpz4=60 p31=2.5m gpnam4=sine.100 5 gpz5=60.75 p29=250u gpnam5=sine.50 A frecuency list FQ2LIST must be defined (list has to be stored in the directory) in which the Caliphatic frequency (43 ppm) is set for first part and the CO frequency (173 ppm) is set for second part.
The standard processing parameters are set to:OBSERVATIONS
Dimension F3 F2 F1 SI 2048 128 512 MC2 - echo/antiecho States-TPPI WDW QSINE QSINE QSINE Offset 2 2 2 ME_mod no LPfc LPfc NCOEF 0 32 32
OTHER RELATED VERSIONS
List of Related Experiments:
Double-Resonance 1H-13C experiments
Double-Resonance 1H-15N experiments
Triple-Resonance 1H-13C-15N backbone experiments
Triple-Resonance 1H-13C-15N sidechain experiments