2D DEPT-based HETCOR experiment
(magnitude mode)

PRELIMINARY SET-UP
  • PARAMETER SET
    a 2D 13C-1H DEPT-based HETCOR spectrum with . Change the pulse program to hxdeptqf
    MODIFY SPECIFIC PARAMETERS
    HETCOR-based

    The evolution period is optimized to d2=3.45 ms (1/2J) , corresponding to 1JCH=145 Hz. and the final editing high power p0 pulse applied from the f1 channel is set to 45, 90 or 135 degree pulse as a function of carbon multiplicity:

    Finally, set the number of scans as a function of the sample concentration (by default, ns 8 and ds 4).

    ACQUISITION
    PROCESSING
    PLOT
    OBSERVATIONS

    (au_getlxhco)) (proc_2dpl).

    OTHER RELATED VERSIONS

    List of Related Experiments:

    Phase-cycle 2D 13C-detected experiments