PARAMETER SET
a 2D 13C-1H DEPT-based HETCOR spectrum with . Change the pulse program to hxdeptqfMODIFY SPECIFIC PARAMETERS
HETCOR-basedACQUISITION
The evolution period is optimized to d2=3.45 ms (1/2J) , corresponding to 1JCH=145 Hz. and the final editing high power p0 pulse applied from the f1 channel is set to 45, 90 or 135 degree pulse as a function of carbon multiplicity:
- 45 degree - all positive
- 90 degree - XH only
- 135 degree - XH, XH3 positive, XH2 negative
Finally, set the number of scans as a function of the sample concentration (by default, ns 8 and ds 4).
PROCESSING
PLOT
OBSERVATIONS
OTHER RELATED VERSIONS(au_getlxhco)) (proc_2dpl).
List of Related Experiments:
Phase-cycle 2D 13C-detected experiments