2D COSY-DQF experiment
(magnitude mode)

PRELIMINARY SET-UP
  • PARAMETER SET
    a 2D 1H-1H COSY-DQF spectrum with rpar COSY45SW all and change the pulse program to cosydfqf
    MODIFY SPECIFIC PARAMETERS
    It is generally recommended that COSY experiments be run without sample spinning.

    Finally, set the number of scans as a function of the sample concentration (by default, ns 8 and ds 4).

    ACQUISITION
    PROCESSING
    PLOT
    OBSERVATIONS

    (au_getlcosy)) (proc_2dsym).

    OTHER RELATED VERSIONS
     
    The same scheme can be used to record 2D COSY experiments with a multiple-quantum filter. For instance, after reading the standard parameter set, a 2D COSY-TQF (triple-quantum filtered) experiment can be run changing the pulse program cosyqftf and the minimum number of scans (ns=12 and ds=24). All other experimental set-up is performed as described above. Such new parameters can be also stored to be used later.

    List of Related Experiments:

    Phase-cycle COSY experiments
    Gradient-enhanced COSY experiments