The calibration of the gradient strength is usually performed by:
  • Measuring the diffusion rate for the residual proton water line in the calibration sample at 25C, and back calculating G. This assumes that D(HDO) in D2O is 1.90*10-5 cm2/s.
  • Acquiring an spin-echo FID with the z-axis gradient on during acquisition. This yields a spatial profile of the sample, which is function of the sample height and the gradient strength.