In special situations it can be advisable to perform 1H acquisition with selective X-decoupling in inverse 1H-X detected experiments of natural abundance and isotopically enriched compounds. Thus during 1H acquisition, selective decoupling of X (13C, 15N ...) is performed from an independent channel using a low-power continuous-wave irradiation.REQUIREMENTS
Easy implementation on AVANCE spectrometers.EXPERIMENTAL DETAILS
For more experimental details see Tutorials: Calibration pulses.SYNTAX
The standard way to implement a selective heteronuclear decoupling is:RELATED TOPICSin which:
...
2 d1 do
...
d12 pl14:f2
go=2 ph31 cw:f2
...
d12 do:f2
...
- go starts data acquisition during an acq period using the acquisition parameters previously defined in eda.
- Broadband heterodecoupling is performed at low power (pl14) from the f2 channel using the cw:f2 command.
- The number after go (=2) defines the loop to be executed. The number of times this loop is repeated is governed by the number of scans (ns).
- ph31 is the receiver phase specified at the end of pulse program and must be cycled in accordance to all pulses involved in the pulse sequence.
- Selective heterodecoupling is stopped using the do command.