REQUIREMENTS
Easy implementation on AVANCE spectrometers.EXPERIMENTAL DETAILS
The most important parameter to optimize is the mixing time (d8).SYNTAX
The standard way to implement a NOESY block with water flip-back is:RELATED TOPICSin which:
...
"DELTA=d8-p16-d16-p11-d12-54u"
...
p1 ph1
DELTA
50u UNBLKGRAD
p16:gp1
d16 pl0:f1
(p11:sp1 ph2:r):f1
4u
d12 pl1:f1
p1 ph3
...
- p1 is the 90 1H pulse (in usec) applied at high power level pl1 from the f1 channel.
- d8 is the mixing NOESY time (in seconds). The delay DELTA is automatically calculated from the expression "DELTA=d8-p16-d16-p11-d12-54u"
- gpz1=50 stands for a purgued gradient of duration p16 (about 1ms)
- Parameters for the water-selective 90 pulse: p11 is the duration, sp1 is the power level, spnam1 is the shape, and phcor2 is the phase difference between hard and soft pulses.
- All phases may be specified at the end of pulse program.